Please use this identifier to cite or link to this item: https://repositorio.ufu.br/handle/123456789/39074
ORCID:  http://orcid.org/0009-0006-9271-6998
Document type: Dissertação
Access type: Acesso Aberto
Title: Avaliação de doses envolvidas em radiocirurgias estereotáxicas utilizando simulação computacional
Alternate title (s): Evaluation of doses involved in stereotactic radiosurgery using computer simulation
Author: Ribeiro, Victor Luiz Mendes
First Advisor: Neves, Lucio Pereira
First member of the Committee: Neves, Lucio Pereira
Second member of the Committee: Belinato, Walmir
Third member of the Committee: Santos, William De Souza
Summary: Radiocirurgia Estereotáxica é um seguimento dentro da radioterapia, onde pacientes são submetidos a altas doses de radicação para erradicar o diminuir doenças onde a cirurgia convencional não é indicada. O objetivo principal do estudo foi avaliar as doses de radiação nos órgãos periféricos de uma paciente gestante, utilizando simulações computacionais empregando o código MCNP6.2. O estudo é inovador, pois combina simulações computacionais com análises detalhadas da distribuição da dose em diferentes tecidos e órgãos do corpo humano. Os resultados obtidos mostram uma considerável variabilidade na dose de radiação entre os órgãos, dependendo da configuração do tratamento em relação à dose de prescrição e das características do tecido irradiado. Empregando o método de Monte Carlo, para prescrições de 15, 18 e 24 Gy, as doses no feto foram de 0,49 cGy, 0,59 cGy e 0,78 cGy respectivamente. Estes valores são para a soma de todos os órgãos do feto.
Abstract: Stereotactic radiosurgery is a follow-up within radiotherapy, where patients are subjected to high doses of radiation to eradicate or reduce diseases where conventional surgery is not indicated. The main objective of the study was to evaluate the radiation doses in the peripheral organs of a pregnant patient, using computational simulations using the MCNP6.2 code. The study is innovative, as it combines computer simulations with detailed analyzes of the dose distribution in different tissues and organs of the human body. The results obtained show considerable variability in the radiation dose between organs, depending on the treatment configuration in relation to the prescription dose and the characteristics of the irradiated tissue. Using the Monte Carlo method, for prescriptions of 15, 18 and 24 Gy, the fetal doses were 0.49 cGy, 0.59 cGy and 0.78 cGy respectively. These values are for the sum of all organs in the fetus
Keywords: Simulação de Monte Carlo
Gestante
Radiocirurgia estereotáxica
Doses no feto
Monte Carlo Simulation
Stereotactic radiosurgery
Pregnant
Doses in fetuses
Area (s) of CNPq: CNPQ::ENGENHARIAS::ENGENHARIA BIOMEDICA
Subject: Engenharia biomédica
Grávidas - Radioterapia
Grávidas - Tratamento intensivo neonatal
Feto - Efeito das drogas
Language: por
Country: Brasil
Publisher: Universidade Federal de Uberlândia
Program: Programa de Pós-graduação em Engenharia Biomédica
Quote: RIBEIRO, Victor Luiz Mendes. Avaliação de doses envolvidas em radiocirurgias estereotáxicas utilizando simulação computacional. 2023. Dissertação (Mestrado Engenharia Biomédica) - Universidade Federal de Uberlândia, Uberlândia, 2023. DOI http://doi.org/10.14393/ufu.di.2023.449
Document identifier: https://doi.org/10.14393/ufu.di.2023.449
URI: https://repositorio.ufu.br/handle/123456789/39074
Date of defense: 4-Aug-2023
Appears in Collections:DISSERTAÇÃO - Engenharia Biomédica

Files in This Item:
File Description SizeFormat 
AvaliaçaoDosesDurante.pdfDissertação2.21 MBAdobe PDFThumbnail
View/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.