Título: Avaliação da dose fetal em gestantes com uso de protetores abdominais em tomografia de crânio via método de Monte Carlo
Título (s) alternativo (s): Fetal dose assessment in pregnant women using abdominal shields in cranial tomography via the Monte Carlo method
Autor: Catusso, Leonardo
Primer orientador: Santos, William de Souza
Primer miembro de la banca: Valença, João Vinícius Batista
Segundo miembro de la banca: Jornada, Tiago da Silva
Resumen: Neste trabalho foram realizadas simulações com o código MCNPX 2.7.0, permitindo estimar a dose nas estruturas de interesse dosimétrico do feto, para o protocolo de crânio do exame de tomografia computadorizada ao qual a gestante foi submetida. Os resultados mostraram que a aplicação de um protetor sobre o abdômen da paciente acarretou uma redução de dose maior que 40% para todas as estruturas do feto, para ambos os fabricantes de TC avaliados. Apesar de os equipamentos simulados terem mesma corrente (mA), colimação, pitch e tempo de rotação, possuíam kVp (130 kV para a Siemens e 120 kV para a GE) e espectros distintos. Essa alta efetividade do protetor ocorre pela baixa energia dos fótons que chegam ao feto, que são provenientes do feixe secundário, vindos do espalhamento na paciente e nas próprias estruturas constituintes do cenário. Na gestante não houve diferença na dose efetiva de maneira significativa, somente as estruturas cobertas pelo protetor apresentaram uma redução de dose maior que 10%, e como esperado as maiores doses absorvidas foram nas estruturas que foram submetidas ao feixe primário, que estavam dentro dos objetivos clínicos de imageamento. A paciente considerada possuía 24 semanas de gestação, 63 kg e 1,63 metros de altura, sendo esse simulador conhecido como Katja. Para o feto quando comparado ambos os fabricantes o fabricante Siemens depositou uma dose cerca de 3% maior que o fabricante GE. Já para a gestante o equipamento GE depositou mais dose, quando se compara a dose efetiva, sendo esse panorama repetido tanto para o caso com protetor quando para o caso sem protetor. Os erros estatísticos da simulação se encontraram dentro do aceitável (menor que 10%) e foram maiores em estruturas que estavam distantes do feixe primário e que possuíam menos voxels.
Abstract: In this study, simulations were performed using the MCNPX 2.7.0 code, allowing the estimation of the dose in the fetal dosimetric structures of interest for the cranial protocol of the computed tomography (CT) exam to which the pregnant patient was subjected. The results showed that the use of a shield over the patient's abdomen resulted in a dose reduction of more than 40% for all fetal structures, for both CT manufacturers evaluated. Although the simulated equipment had the same current (mA), collimation, pitch, and rotation time, they had different kVp values (130 kV for Siemens and 120 kV for GE) and distinct spectra. This high shield effectiveness occurs due to the low energy of the photons reaching the fetus, which originate from the secondary beam, resulting from scattering within the patient and the structures constituting the scenario. For the pregnant patient, there was no significant difference in the effective dose; only the structures covered by the shield showed a dose reduction of more than 10%. As expected, the highest absorbed doses were in structures exposed to the primary beam, which were within the clinical imaging objectives. The simulated patient was 24 weeks pregnant, weighed 63 kg, and was 1.63 meters tall, represented by the Katja phantom. For the fetus, when comparing both manufacturers, Siemens equipment deposited a dose approximately 3% higher than GE equipment. However, for the pregnant patient, the GE equipment deposited a higher dose when comparing the effective dose. This trend was observed in both the shielded and unshielded cases. The statistical errors of the simulation were within acceptable limits (less than 10%) and were higher in structures far from the primary beam and those with fewer voxels.
Palabras clave: Radiodiagnóstico
Tomografia computadorizada
Simulação Monte Carlo
Dose fetal
Simulador antropomórfico virtual
Área (s) del CNPq: CNPQ::CIENCIAS EXATAS E DA TERRA::FISICA
Tema: Engenharia Biomédica
Tomografia computadorizada espiral
Diagnóstico radioscópico
Idioma: por
País: Brasil
Editora: Universidade Federal de Uberlândia
Programa: Programa de Pós-graduação em Engenharia Biomédica
Cita: CATUSSO, Leonardo. Avaliação da dose fetal em gestantes com uso de protetores abdominais em tomografia de crânio via método de Monte Carlo. 2025. 64 f. Dissertação (Mestrado em Engenharia biomédica) - Universidade Federal de Uberlândia, Uberlândia, 2025. DOI http://doi.org/10.14393/ufu.di.2025.19.
Identificador del documento: https://doi.org/10.14393/ufu.di.2025.19
URI: https://repositorio.ufu.br/handle/123456789/44930
Fecha de defensa: 13-ene-2025
Objetivos de Desarrollo Sostenible (ODS): ODS::ODS 11. Cidades e comunidades sustentáveis - Tornar as cidades e os assentamentos humanos inclusivos, seguros, resilientes e sustentáveis.
Aparece en las colecciones:DISSERTAÇÃO - Engenharia Biomédica

Ficheros en este ítem:
Fichero Descripción TamañoFormato 
downloadfile.bin.pdfAutorização de publicação1.4 MBAdobe PDFVista previa
Visualizar/Abrir
AvaliaçãoDoseFetal.pdfDissertação1.65 MBAdobe PDFVista previa
Visualizar/Abrir


Los ítems de DSpace están protegidos por copyright, con todos los derechos reservados, a menos que se indique lo contrario.