Título: Avaliação de implantes sem fios de sutura para o reparo alternativo de tendões do manguito rotador
Título (s) alternativo (s): Avaliação de implantes sem fios de sutura para o reparo alternativo de tendões do manguito rotador
Autor: Guimarães, Marcília Valéria
Primer orientador: Araújo, Cleudmar Amaral de
Primer coorientador: Campos, Dagoberto de Oliveira
Primer miembro de la banca: Mello , José Daniel Biasoli de
Segundo miembro de la banca: Gonçalves, Rogério Sales
Tercer miembro de la banca: Zavaglia, Cecília Amélia de Carvalho
Cuarto miembro de la banca: Barbosa, Maria Helena
Resumen: Lesões do manguito rotador são desordens patológicas que provocam dor e disfunção do ombro. O reparo cirúrgico destas lesões é feito comumente usando âncoras de sutura por meio da artroscopia. Porém, as âncoras possuem desvantagens como a passagem dos fios de sutura pelo tendão e a amarração dos nós que tornam a cirurgia mais complexa. Para contornar estes problemas, este trabalho propôs desenvolver e otimizar implantes que não usam fios de sutura. O comportamento mecânico dos implantes foi avaliado usando ensaios experimentais e modelagens por elementos finitos. Como resultado, três novos implantes em forma de grampo foram propostos. Estes implantes foram fabricados em liga de titânio por manufatura aditiva e nomeados de implante tipo 1 (duas hastes e cabeça retangular), tipo 2 (duas hastes inclinadas e cabeça retangular alargada) e tipo 3 (três hastes e cabeça circular). Ensaios mecânicos foram conduzidos usando uma fita de poliéster passada pela cabeça dos implantes para tracioná-los e um bloco de espuma rígida com propriedades mecânicas próximas às do osso trabecular para inserção dos implantes. A força média de inserção de cada implante foi estimada por meio de um aparato construído para esse procedimento e conectado a um sistema de aquisição de dados. A resistência mecânica e o deslocamento dos implantes foram avaliados em testes de arrancamento a uma velocidade constante. A análise da fixação dos implantes foi feita por meio de modelos tridimensionais de elementos finitos. Os resultados dos ensaios mostraram que durante a inserção dois implantes falharam, um do tipo 1 e outro do tipo 3. Na inserção, o implante tipo 3 necessitou de uma força média maior para ser inserido do que os implantes tipo 1 e tipo 2. No arrancamento, a resistência de fixação do implante tipo 2 foi superior à obtida para os implantes tipo 1 e tipo 3. E o deslocamento até a falha do implante tipo 2 foi menor que o dos implantes tipo 1 e tipo 3. Nas análises por elementos finitos, o modelo de reparo com o implante tipo 2 apresentou na região distal do tendão os menores níveis de deslocamento resultante e tensão de von Mises. A partir dos resultados concluiu-se que o implante tipo 2 apresentou características geométricas que podem contribuir para uma melhor resistência da fixação e estabilidade do tendão ao osso em reparos do manguito rotador.
Abstract: Rotator cuff tears are pathological disorders that generate shoulder pain and disability. The repair is made by arthroscopic surgery using suture anchors. However, the suture anchors have disadvantages like guiding sutures through tendon and tying knots. To solve those problems, the aim of this study is developing and optimizing sutureless implants. The mechanical behavior of implants was evaluated using experimental testing and finite element analysis. As result, two three staple-shaped implants was proposed. The implants were manufactured in titanium alloy by additive manufacturing and named of implant type 1 (two legs and rectangular head), type 2 (two legs and enlarged rectangular head) and type 3 (three legs and circular head). Experimental testing was performed with a polyester ribbon passed through the head of implants to pull them out and a rigid polyurethane foam block with close related biomechanical properties to the trabecular bone. The insertion average load of each implant was estimated by using a device building for that procedure and connected to a data acquisition system. The mechanical strength and displacement of implants was evaluated using pull out testing at a constant rate. The analysis of implant fixation was performed by using tridimensional finite element models. The results of experimental testing showed that two implants had failed during the insertion procedure: one of type 1 and one of type 3. The implant type 3 required a higher average load to be inserted than the others. In the pullout test, the fixation strength of implant type 2 was higher than type 1 and 3. And the displacement until the failure of the implant type 2 was lower than the other implants. In the finite element analysis, the rotator cuff repair model of implant type 2 showed lower levels of displacement and von Mises stress. From these results it was concluded that the type 2 implant showed geometrical characteristics that can contribue to a better fixation resistance and stability of bone tendon in rotator cuff repairs.
Palabras clave: Lesões
Engenharia Mecânica
Elementos finitos
Tendões - Implantes
Manguito Rotador
Ombro - Lesões
Novos implantes sem fios de sutura
Ensaios experimentais
Suturas
Rotator cuff
Tears
Finite element analysis
New sutureless implants
Experimental testing
Área (s) del CNPq: CNPQ::ENGENHARIAS::ENGENHARIA MECANICA::MECANICA DOS SOLIDOS::MECANICA DOS CORPOS SOLIDOS, ELASTICOS E PLASTICOS
Idioma: por
País: Brasil
Editora: Universidade Federal de Uberlândia
Programa: Programa de Pós-graduação em Engenharia Mecânica
Cita: GUIMARÃES,Marcília Valéria. Avaliação de implantes sem fios de sutura para o reparo alternativo de tendões do manguito rotador. 2018. 182 f. Tese (Doutorado em Engenharia Mecânica) - Universidade Federal de Uberlândia, Uberlândia, Uberlândia, 2018. DOI http://dx.doi.org/10.14393/ufu.te.2019.7.
Identificador del documento: http://dx.doi.org/10.14393/ufu.te.2019.7
URI: https://repositorio.ufu.br/handle/123456789/27044
Fecha de defensa: 31-jul-2018
Aparece en las colecciones:TESE - Engenharia Mecânica

Ficheros en este ítem:
Fichero Descripción TamañoFormato 
AvaliacaoImplantes.pdf8.71 MBAdobe PDFVista previa
Visualizar/Abrir


Los ítems de DSpace están protegidos por copyright, con todos los derechos reservados, a menos que se indique lo contrario.