Please use this identifier to cite or link to this item:
https://repositorio.ufu.br/handle/123456789/20769
Document type: | Dissertação |
Access type: | Acesso Aberto |
Title: | Avaliação do desempenho de unidade eletrocirúrgica |
Alternate title (s): | Evaluation of the performance of electrosurgical unit |
Author: | Silva, Marco Tullio Alves |
First Advisor: | Arencibia, Rosenda Valdés |
First coorientator: | Milagre, Selma Terezinha |
First member of the Committee: | Franco, Sinésio Domingues |
Second member of the Committee: | Moraes, Raimes |
Summary: | Este trabalho tem como objetivo avaliar o desempenho metrológico de 15 unidades eletrocirúrgicas de dois fabricantes e três modelos diferentes quando submetidos a variações de temperatura ambiente. Para tanto foram considerados cinco níveis do fator temperatura e três pontos da faixa nominal das potências de corte e de coagulação. Os ensaios foram realizados acoplando a unidade eletrocirúrgica objeto de avaliação a um analisador de potência. O teste de Tukey foi aplicado visando efetuar múltiplas comparações entre os níveis do fator temperatura. Os parâmetros metrológicos incerteza de medição, erro, repetibilidade e erro máximo foram estimados para todos as unidades avaliadas. A incerteza associada à medição foi avaliada de acordo com as recomendações do Guia para a Expressão da Incerteza de Medição (GUM). A análise de variância (ANOVA), o teste de Tukey e o boxplot indicaram efeitos estatisticamente significativos do fator temperatura nos valores de potência de corte e coagulação avaliados. A incerteza expandida associada aos valores de potência de corte e de coagulação aumentou de forma significativa na medida em que estas se aproximaram do limite superior da faixa nominal. Foram observadas diferenças significativas entre as unidades eletrocirúrgicas dos fabricantes A e B sobre tudo para menores valores de potência. Neste sentido as unidades do fabricante A apresentaram desempenho superior. Nas potências de corte 50 W, 150 W e 300 W, observou-se que 33 %, 87 % e 100 %, respectivamente, das unidades avaliadas apresentaram valores de erro máximo maiores que 5 W e portanto não atendem ao critério especificado pelo fabricante. Para as potências de coagulação, 30 W, 80 W e 120 W estas porcentagens foram respectivamente de 0, 53 % e 60 %. As unidades eletrocirúrgicas avaliadas apresentam um desempenho que podem comprometer o sucesso do processo cirúrgico. |
Abstract: | The objective of this work is to evaluate the metrological performance of 15 electrosurgical unit from two manufacturers and three different models when subjected to ambient temperature variations. For this, five levels of the temperature factor and three points of the nominal range of the cutting and coagulation powers were considered. The tests were performed by coupling the evaluation electrosurgical unit to a power analyzer. The Tukey test was applied in order to make multiple comparisons between the levels of the temperature factor. The metrological parameters measurement uncertainty, error, repeatability and maximum error were estimated for all evaluated electrosurgical unit. The uncertainty associated with the measurement was evaluated according to the recommendations of the Guide to the Expression of Uncertainty in Measurement (GUM). Analysis of variance (ANOVA), Tukey test and boxplot indicated significant statistically effects of the temperature factor on the cutting and coagulation power values. The expanded uncertainty associated with cutting and coagulation power values increased significantly as they approached the upper limit of the nominal range. Significant differences were observed between the scalpels of manufacturers A and B mainly for lower power values. The scalpels of the manufacturer A presented superior performance. At 50 W, 150 W and 300 W cutting powers, it was observed that 33 %, 87 % and 100 %, respectively, of the scalpels had maximum error values greater than 5 W and therefore they did not meet the criterion specified by the manufacturer (maximum error must be less than 5 W). For the coagulation powers, 30 W, 80 W and 120 W these percentages were respectively 0, 53 % and 60 %. The electrosurgical unit evaluated have a metrological performance that can compromise the success of the surgical process. |
Keywords: | Electrosurgical unit Cutting power Coagulation power Room temperature Measurement uncertainty Unidade Eletrocirúrgica Potência de corte Potência de coagulação Temperatura ambiente Incerteza de medição Engenharia mecânica Instrumentos de medição Tecnologia em serviços de saúde Tecnologia médica |
Area (s) of CNPq: | CNPQ::ENGENHARIAS::ENGENHARIA MECANICA::PROCESSOS DE FABRICACAO |
Language: | por |
Country: | Brasil |
Publisher: | Universidade Federal de Uberlândia |
Program: | Programa de Pós-graduação em Engenharia Mecânica |
Quote: | SILVA, Marco Tullio Alves. Avaliação do desempenho de unidade eletrocirúrgica. 2017. 195 f. Dissertação (Mestrado em Engenharia Mecânica) - Universidade Federal de Uberlândia, Uberlândia, 2017. |
Document identifier: | http://dx.doi.org/10.14393/ufu.di.2018.165 |
URI: | https://repositorio.ufu.br/handle/123456789/20769 |
Date of defense: | 28-Jul-2017 |
Appears in Collections: | DISSERTAÇÃO - Engenharia Mecânica |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
AvaliacaoDesempenhoUnidade.pdf | Dissertação | 13.17 MB | Adobe PDF | View/Open |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.