Título: Avaliação dos fatores que afetam a análise de imagens de raio-X para diagnóstico de cifose, escoliose e osteoporose
Título(s) alternativo(s): Evauation of factors affecting the analysis of X-ray images for diagnosis of kyphosis, escoliosis and osteoporosis
Autor(es): Palis, Angélica de Andrade
Primeiro orientador: Arencibia, Rosenda Valdés
Primeiro coorientador: Milagre, Selma Terezinha
Primeiro membro da banca: Reis, Ruham Pablo
Segundo membro da banca: Moraes, Raimes
Resumo: O presente trabalho tem como objetivo avaliar o efeito do fator Operador nos valores de ângulo de Cobb obtidos em imagens de raio-X de pacientes que foram diagnosticados com escoliose, cifose e osteoporose, bem como propor procedimentos para avaliação da incerteza de medição deste mensurando levando em conta cada sistema de medição utilizado. Para tanto o ângulo de Cobb foi medido por três Operadores em imagens analógicas, digitais e computadorizadas no diagnóstico de escoliose e cifose. Dois sistemas de medição foram utilizados para efetuar as medições, o transferidor analógico (imagens analógicas) e o software Kinovea (imagens digitais e computadorizadas). Para o diagnóstico de osteoporose foi efetuada a medição apenas em imagens digitais utilizando os softwares Prodigy Lunar – BMD e Kinovea. A análise dos resultados foi efetuada aplicando análise de variância (ANOVA) por meio do software Statistica R. A incerteza de medição foi estimada seguindo as recomendações do método GUM proposto no JCGM 101. Os resultados indicaram que o Operador 1 “Experiente” apresentou a melhor repetibilidade dos resultados e consequentemente a menor incerteza expandida. As imagens com maior curvatura na coluna vertebral apresentaram pior repetibilidade dos valores médios de ângulo de Cobb e consequentemente valores maiores de incerteza expandida, o que demonstra a dificuldade de identificar as vértebras limites por parte dos Operadores quando os ângulos são maiores. Os valores de incerteza expandida (95%) associados ao ângulo de Cobb para diagnóstico de osteoporose podem ser da mesma ordem do valor do mensurando em alguns casos.
Abstract: The present study aims to evaluate the effect of the factor Operator on Cobb angle values obtained on X-ray images of patients who were diagnosed with scoliosis, kyphosis and osteoporosis, as well as to propose procedures to evaluate the measurement uncertainty of this measurand taking into account each measuring system used. For this, the Cobb angle was measured by three Operators in analog, digital and computerized images in the diagnosis of scoliosis and kyphosis. Two measurement systems were used to carry out the measurements, the analog transponder (analogue images) and the Kinovea software (digital and computerized images). For the diagnosis of osteoporosis, measurements were made only on digital images using Prodigy Lunar software - BMD and Kinovea. The analysis of the results was carried out by applying analysis of variance (ANOVA) using Statistica R. The measurement uncertainty was estimated following the recommendations of the GUM method proposed in JCGM 101. The results indicated that Operator 1 "Experienced" presented the best repeatability of the results and consequently the lower uncertainty. The images with greater curvature in the vertebral column presented worse repeatability of the average values of Cobb angle and consequently higher values of expanded uncertainty, which demonstrates the difficulty to identify the vertebrae limits by the Operators when the angles are larger. The expanded uncertainty values (95%) associated with the Cobb angle for diagnosis of osteoporosis may be of the same order as the value of the measurand in some cases.
Palavras-chave: Escoliose
Cifose
Osteoporose
Raio-X
Anova
Scoliosis
Kyphosis
Osteoporosis
X-ray images
Anova
Área(s) do CNPq: CNPQ::ENGENHARIAS
Idioma: por
País: Brasil
Editora: Universidade Federal de Uberlândia
Programa: Programa de Pós-graduação em Engenharia Mecânica
Referência: PALIS, A. A. Avaliação dos fatores que afetam a análise de imagens de raio-X para diagnóstico de cifose, escoliose e osteoporose. 2017. 113 f. Dissertação de Mestrado, Universidade Federal de Uberlândia, Uberlândia.
Identificador do documento: http://dx.doi.org/10.14393/ufu.di.2018.41
URI: https://repositorio.ufu.br/handle/123456789/20666
Data de defesa: 28-Abr-2017
Aparece nas coleções:DISSERTAÇÃO - Engenharia Mecânica